[摘 要]
Discovery XR656目前是GE DR的最高端产品。Discovery XR656融合了当代先进的平板技术和GE的Discovery平台。本文
简述了GE Discovery XR656 DR系统硬件及全景成像功能,并分享了一例全景拼接功能故障的分析及排除过程,总结设备维修
及日常管理的发展方向。
[关键词]
医疗设备;全景成像;故障检修;放射控制接口模块
GE Discovery XR656 DR Hardware Overview and Panorama Imaging Function
Fault Analysis and Troubleshooting
ZHANG Xingkun
1
, LIU Hongguang
1
, LI Weikai
2
1. Department of Medical Technology, Qingdao Central Hospital, Qingdao Shandong 266042, China;
2. Department of Medical Imaging, Qingdao Women and Children’s Hospital, Qingdao Shandong 266034, China
Abstract:
Discovery XR656 is currently the most high-end product of GE DR. Discovery XR656 combines modern advanced
Flash Pad technology with GE’s Discovery platform. We briefly described the hardware and panoramic imaging functions of GE
Discovery XR656 DR system, shared a case of fault analysis and troubleshooting of panoramic stitching function, and summarized
the development direction of equipment maintenance and daily management in this paper.
Key words:
medical equipment; panorama imaging function; troubleshooting; radiation control interface module
R197.3
[文献标识码]
B
[中图分类号]
doi:
10.3969/.1674-1633.2019.10.049
[文章编号]
1674-1633(2019)10-0189-04
引言
Discovery XR656 DR基于全新Discovery科研功能平
台、全新数字无线UWB(Ultra-Wideband超宽带)、平板
探测器Flash Pad,可实现优越的量子捕获效率和信噪比,
并实现了独有Volume RAD容积成像、双能成像、全景成
像、组织均衡等功能,突破了前后组织信息重叠的二维投
影影像,在呼吸、骨骼、泌尿系统的影像学诊断上颇具优势,
可大大提高传统放射科的工作效率,充分展现了放射学全
数字化无限发展的前景。该设备自动化水平极高,且存在
大量复杂的系统,需要用到计算机技术、电子信息技术等
不同方面的知识,对维修技术人员自身的工作水平和能力
提出更高的要求
[1-4]
。本文主要简介GE Discovery XR656
DR系统硬件及全景拼接功能,并分享了一例全景拼接功能
故障的分析及排除过程,供同行参考。
系统”。现在普遍应用的DR主要是采用平板探测口对X
线产生的图像信号进行扫描和直接读出,成像原理是先将
X线信号转变为可见光通过利用硒层或碘化铯-光电2极
管组成的藻膜层进行聚集,由专门的读出电路直接读出直
接把X线光子转换成模拟电压供数字化后送计算机系统
进行处理,所有过程全部在平板探测器内完成。Discovery
XR656 DR系统由多个主要组件和子组件构成。如图1所
示系统主要组件有采集工作站、射线管高位托架(OTS)、
球管、系统柜(含高压发生器)、数字化探测器(两套GE
Flash Pad)、数字化扫描床和墙架。
1 Discovery XR656 DR系统硬件及全景拼接功能概述
1.1 Discovery XR656 DR系统硬件概述
DR是Digital radiography 的简称,即“数字放射成像
收稿日期:2018-12-17
作者邮箱:zxk1970@
图1 Discovery XR656 DR主要系统组件
采集工作站配备有专用计算机和图像数据库及放射控
制接口模块(Radiation Control Interface Module,RCIM)
中国医疗设备 2019年第34卷 10期
VOL.34 No.10
189
设备维修
DEVICE MAINTENANCE
和控制台手控开关,使用控制台手控开关可进行曝光,
RCIM主要用于控制系统的电源和复位功能。采集工作站
主要进行图像采集,图像显示和操作并通过DICOM标准
传输到其他工作站。射线管高位托架(OTS)是用来支撑
X射线管和OTS用户界面的定位设备。OTS组件可以在多
个方向上移动,以便在患者检查过程中正确进行装置定位。
系统柜中主要包含高压油箱、电源分配单元及各种通讯板
等。主要负责控制台、射线管高位托架OTS、扫描床、墙
架的电源分配及其相互间通讯。数字化探测器构造采用碳
纤维主要部件有面板和电子装置,负责原始数据的采集并
模数转换为数字图像数据,然后内部网络将此数据传输到
系统工作站,其他数字化墙架及扫描床用于完成各种放射
摄影程序的患者体位摆放与数字化探测器位置移动
[5-8]
。
1.2 全景拼接功能概述
全景拼接功能是一个高级应用程序,可将一个患者相
邻部位的两次甚至三次四次不同曝光区域的影像进行自动
拼接,从而突破了探测器尺寸的限制,可获得胸腹部、上
下肢等的整体影像。主要用于骨关节系统疾病等方面,能
可靠、准确地显示病变局部改变和全脊柱/全肢体的整体
受力状态,为临床诊断、术前方案的制定和术后疗效评估
等提供可靠依据
[9]
。自动图像粘贴可以在墙架和扫描床接
收器上使用。DR拼接图像的原始图像采集方式包括两种:
一种是球管与DR平板探测器做垂直上下的同步移动,另
一种是如图2a所示,X线球管组件相对静止在一个感兴趣
区中心位置,当DR平板探测器按箭头图示方向在做垂直
运动的同时,X线球管配合按箭头所示做上下转动角度。
系统先按顺序采集2~5次曝光采集到每个部位的影像,然
后再用数字方式将它们粘贴在一起,形成一张图像,采集
后的子图像会出现在机器浏览器屏幕上,而全景拼接的图
像会在大约10 s后出现
[10-11]
。
图2 墙架侧全景拼接采集示意图
注:a. 全景拼接硬件运动示意图;b. 每个部位影像;c. 全景
拼接图像。
2 Discovery XR656 DR设备全景拼接故障及其维修
方法
2.1 故障描述
设备开机后,设备信息提示栏偶发提示“system
control communication interrupted. Could be possible cable
disconnection. Call service if problem persists.”正常DR图
像功能正常,但频发全景拼接功能异常,提示“Exposure
中国医疗设备 2019年第34卷 10期
VOL.34 No.10
was prematurely ended. Image quality may be affected”等信
息,如图3所示,将图3a测试物品置于全景拼接患者固定
架上,反复功能测试,连续曝光频发中断(如图3b浏览器
图像缺失),全景拼接失败,无法正常连续曝光采集到如图
3c所示全景拼接成功获取图像,多次反复曝光捕捉故障点,
发现检查床及墙架侧进行全景扫描故障都会发生,并且与
球管及探测器的位置无关。
图3 全景拼接功能故障演示
注:a. 测试物品;b. 浏览器图像缺失;c. 全景拼接成功获取
图像。
2.2 故障分析
系统控制台信息提示栏偶发系统控制通讯中断提示,
但正常DR检查功能正常,说明正常DR曝光时,数据是
正常可以从平板探测器传至控制台,通讯正常,考虑到正
常DR曝光时间周期极短,说明系统内部通讯状态不稳定
而并未完全中断。而结合全景拼接功能故障时系统控制台
信息提示栏提示的曝光意外终止,综合分析全景拼接功能
故障可能的原因可以从以下几个方面去寻找:① 线路物理
连接问题导致的通讯故障,设备开机后,报通讯错误,最
大的可能是通讯连接线有松动,或者是插口处长时间使用
出现氧化,导致接触不良的虚接现象;② 软件问题,曝光
协议,应用程序故障,系统软件问题等;③ 相关硬件问题,
通信处理板、曝光控制盒等隐性硬件故障。
本着从低值到高值原则。首先检查负责通信的连接线
路是否正常及连接线接口是否因为机械运动等原因导致有
松动,或者是插口处长时间使用出现氧化,导致接触不良
的虚接等现象,其次是相关通讯板、电源板等模块的输入
输出特性,再次是排除曝光协议,应用程序故障,系统软
件等软件问题,最后排除通信处理板、曝光控制盒等隐性
硬件故障问题
[12-13]
。
2.3 故障处理
2.3.1 基本物理线路故障排查
关机断电状态下,进行基本物理线路故障排查,主要
发现以下异常点,第一处为如图4a所示,打开墙架前盖板
发现一通信线缆机械磨损严重,线标为“MS020021 FROM
WS A1J1 TO SKL(SYSTEMS CABINET)A25J14”,近墙架
端一处通信接口机械松动严重,使用扎带把线缆重新加固,
使其远离伸展臂运动轨道,并加固通信接口。第二处为如
图4b 所示系统柜内一通信接口松动,予以加固,关机重启,
予以全景拼接故障测试,故障依旧。
190
图4 物理线路异常点
注:a. 墙架物理线路异常点;b. 系统柜物理线路异常点。
2.3.2 软件故障排查
本着从简到繁原则,可先恢复曝光协议,若故障依旧
可重装系统及应用程序软件,若重装系统及应用程序软件
仍不能排除故障,排除系统及应用程序软件问题。当然在
软件故障排除的过程中需要注意的是在曝光协议及系统应
用程序进行还原前,应首先要确保所备份的最近日期的还
原协议及系统应用程序数据完好性,并且最好在还原前做
好备份,以保证万一不是软件问题所引起故障的情况时好
将数据及时还原回去,以便于下一步故障的分析判断。
首先是曝光协议故障排除,还原曝光协议,还原之
前确认最近日期的曝光协议管理备份)数据完好并做
好当前日期的曝光协议备份,Service模式下点击进入
System→Preferences→Protocols选项下有曝光协议的还原
与备份选项以供操作,使用最近日期的曝光协议备份数据
进行曝光协议还原,还原完成后,关机重启,予以全景拼
接故障测试,故障依旧,排除曝光协议导致的故障,使用
当前日期的曝光协议备份数据再次进行还原操作,以便机
器状态是故障排除前状态,供下一步的故障排除。
其次是系统及应用程序软件故障的排除,排除掉曝光
协议问题后,只能重装软件尝试。重装软件之前,确认最
近日期的系统备份数据完好并做好当前日期的系统备份,
若判断故障并非系统及应用程序软件故障引起,及时将设
备数据还原至最初故障排除前状态,供下一步的故障排除。
具体重装软件步骤如下。
(1)操作系统软件安装,将linux操作系统软件安装
光盘放入光驱,按住RESET 5秒钟,重启机器,在显示
器开始闪烁的时候,按住F9,之后会出现一个选择界面,
选择Optical Drive,点击回车,当出现Boot的时候,输出
GEHCXR(必须大写),回车,之后会自动完成安装,完成
安装后,取出光盘,系统会自动重启,进入linux系统。
(2)系统应用程序安装system applications software,
将最近日期的系统应用程序软件安装光盘放入光驱,点击
CDROM图标,并且点击并且运行autorun,机器会自动重
启,当机器重启后,再次打开CDROM图标,并再次运行
autorun,在出现的对话框中选择相对应的机型656,系统
会自动完成安装,当完成安装后,机器会重启,重启后取
出光盘。
DEVICE MAINTENANCE
设备维修
(3)重启完成后,输入账户名:insite密码:2getin进
入应用操作平台,点击右上角的工具图标,选择维修图标,
在维修界面选择Utillties并找到System,找到Restore,将
最近日期的系统备份放入光驱,选择从光驱还原,点击开始,
此处要注意,在如图5所示,出现的对话框中,一定要一
个一个的完成还原,此处应注意,在还原下一选项前一定
要确保上一选项的还原进程有“success”提示出现,在完
成所有选项的还原后,点击退出,将光盘取出。
图5 系统还原弹窗
(4)将曝光协议光盘放入光驱,进入参数设置界面,
选择参数-协议-还原,完成协议的还原,还原完成后,
重启设备,测试PACS和WORKLIST的功能正常,全景拼
接功能测试,故障依旧,且通信报错仍存在,此时应将故
障前所做系统备份数据及曝光协议管理备份数据进行还原,
以便机器状态是故障排除前状态,供下一步的故障排除。
2.3.3 硬件故障排查
鉴于“多次反复曝光捕捉故障点,发现检查床及墙架
时进行全景扫描故障都会发生,且尚未发现故障与球管及
探测器的某一具体位置点关联”的故障现象,可重点考虑
关注检查床及墙架共用部分设备部件如高压柜与控制台部
分,结合控制台一直存在通信故障报错,最后综合分析故
障主要涉及硬件部分优先从通信部分排查,而涉及通信的
物理硬件主要有图6所示系统主通信板和图7所示RCIM。
本着由简入繁原则,先从主通信板故障排查,关机状态下,
打开机箱,负责通信的板子在供电柜门上,将通信板外部
的保护外壳拆下, 各连接线予以重新拔插,无松动,开机
观察通信板上常规的指示灯,指示灯的工作状态都正常。
此时在开机上电状态下用数字万用表测量通信板开关电源
供电电压,电压值为DC23 V,调节如图所示开关电源点位
器,将输出调至24 V,关机重启,功能测试,设备故障依
旧,基本排除通讯板故障,最后只能怀疑图7所示RCIM,
检查其工作状态指示灯及供电电源未见异常,考虑到其板
级集成度极高,只能考虑将其更换,更换RCIM后,系统
通信报错提示消失,全景拼接功能多次测试正常,曝光连
续不中断,能够获取采集到图3c图像所示的图像。
中国医疗设备 2019年第34卷 10期
VOL.34 No.10
191
设备维修
DEVICE MAINTENANCE
[参考文献]
[1] 杨小娟,吴国荣.医疗设备维修管理现状及对策探讨[J].中国
医疗器械信息,2017,74(18):152-153.
[2] 范境文.医院医疗设备维修精细化管理的必要性和措施[J].医
疗装备,2017,30(18):73-74.
[3] 辛胜昌.浅论本院医疗设备维修管理中存在的问题及对策[J].
中国医疗器械信息,2017,23(16):149-151.
[4] 陈晓飞,周晟,汪新柱,等.数字X线摄影焦点至探测器
图6 系统主通信板
距离与体模表面入射剂量的实验研究[J].中国临床研
究,2016,29(3):384-385.
[5] 韩丰谈.医学影像设备学[M].北京:人民卫生出版社,2004:458-
图7 放射控制接口模块(RCIM)
464.
[6] 徐江波.GE Definium6000飞天 DR的使用管理与常见故障维
修[J].医疗装备,2018,31(3):151.
[7] 谈伟,张德洲,黄林,等.数字X线摄影的全景拼接技术在骨关
节系统中的应用研究[J].华西医学,2013,(9):1398-1401.
[8] 熊刚,朱俊义,魏瑜栋.新形势下我院医疗设备维修管理模式
改进方法探讨[J].中国医疗设备,2018,33(9):175-177.
[9] 徐跃,梁碧玲.医学影像设备学[M].3版.北京:人民卫生出版
社,2010:257.
[10] 向思伟,郑琨.GE Defnium6000 DR 故障维修两例[J].中国医
疗设备,2018,33(2):100-101.
[11] 刘伟宾,张燕,吴新淮,等.GE Discovery XR650 X线系统
在儿童双下肢不等长中的临床应用价值[J].医疗卫生装
备,2013,34(8):69-70.
[12] 盛况,孔德友,范磊.GE Defnium 6000数字X线机故障维修分
析3例[J].中国医疗设备,2018,33(1):102-103.
设备,2016,31(10):172-173.
械信息,2017,23(19):130-131.
[13] 于金科,祝健.Defnium 6000 DR 故障维修3例[J].中国医疗
[14] 苏健.加强医疗设备维修服务管理的新策略[J].中国医疗器
[15] 王可寿.医院医疗设备维修的精细化管理[J].医疗装备,
2018,31(9):93-94.
C
3 总结
本文简介了GE Discovery XR656 DR系统硬件及全景
成像功能,并分享了一例全景拼接功能故障的故障现象,
并给出了故障分析及排除过程。随着微电子等各种高新技
术在医疗设备中广泛运用与整合,加之医疗设备厂方的技
术垄断与保护,使得大型医疗设备的维修和维护面临严峻
考验,医院工程师传统的维修模式面临巨大挑战。作为院
方工程师,应从设备装机安装调试开始就应积极加强自身
对机器的了解学习,与此同时,当初期设备在保修期内时,
应加强与厂家工程师的沟通与学习,从机器的基本操作使
用到基本结构原理,从巡检保养各项明细到日常常见故障
的分析的思路和排除方法,为以后独立维修打下良好基础。
而在设备的日常管理当中,应加强设备的巡检保养,制定
严格的巡检保养明细和周期,按期执行并落实到每条明细,
将传统的设备故障后维修的维修模式升级为预防性维修,
及时将故障隐患排除,并且在每次故障后及时详细的记录
故障现象、分析解决的全过程,做好每台设备的动态管理
档案,给设备的精细化、信息化管理提供数据支持。各级
医院应巧妙地运用精细化管理模式,高度重视维修人员专
业技能的培养,提高维修设备的信息化,以此来保障医疗
设备能够高效、正常的投入临床运用中
[14-15]
。
稿件作者更正声明
《中国医疗设备》2019年第34卷第09期185-187页《瓦里Trilogy维修故障一例》一文,本人
在发定稿给《中国医疗设备》编辑部时,由于疏忽,出现了外文文字错误。现更正如下:
原文中出现的所有文字“Trllogy”更正为“Trilogy”。特此补充更正,表示歉意。
陆奇鹏
2019年9月24日
192
中国医疗设备 2019年第34卷 10期
VOL.34 No.10
2024年4月6日发(作者:泥致远)
DEVICE MAINTENANCE
设备维修
GE Discovery XR656 DR硬件概述及
全景成像功能故障分析及排除
张行坤
1
,刘红光
1
,李伟凯
2
1. 青岛市中心医院 医技部,山东 青岛 266042;2. 青岛市妇女儿童医院 医学影像科,山东 青岛 266034
[摘 要]
Discovery XR656目前是GE DR的最高端产品。Discovery XR656融合了当代先进的平板技术和GE的Discovery平台。本文
简述了GE Discovery XR656 DR系统硬件及全景成像功能,并分享了一例全景拼接功能故障的分析及排除过程,总结设备维修
及日常管理的发展方向。
[关键词]
医疗设备;全景成像;故障检修;放射控制接口模块
GE Discovery XR656 DR Hardware Overview and Panorama Imaging Function
Fault Analysis and Troubleshooting
ZHANG Xingkun
1
, LIU Hongguang
1
, LI Weikai
2
1. Department of Medical Technology, Qingdao Central Hospital, Qingdao Shandong 266042, China;
2. Department of Medical Imaging, Qingdao Women and Children’s Hospital, Qingdao Shandong 266034, China
Abstract:
Discovery XR656 is currently the most high-end product of GE DR. Discovery XR656 combines modern advanced
Flash Pad technology with GE’s Discovery platform. We briefly described the hardware and panoramic imaging functions of GE
Discovery XR656 DR system, shared a case of fault analysis and troubleshooting of panoramic stitching function, and summarized
the development direction of equipment maintenance and daily management in this paper.
Key words:
medical equipment; panorama imaging function; troubleshooting; radiation control interface module
R197.3
[文献标识码]
B
[中图分类号]
doi:
10.3969/.1674-1633.2019.10.049
[文章编号]
1674-1633(2019)10-0189-04
引言
Discovery XR656 DR基于全新Discovery科研功能平
台、全新数字无线UWB(Ultra-Wideband超宽带)、平板
探测器Flash Pad,可实现优越的量子捕获效率和信噪比,
并实现了独有Volume RAD容积成像、双能成像、全景成
像、组织均衡等功能,突破了前后组织信息重叠的二维投
影影像,在呼吸、骨骼、泌尿系统的影像学诊断上颇具优势,
可大大提高传统放射科的工作效率,充分展现了放射学全
数字化无限发展的前景。该设备自动化水平极高,且存在
大量复杂的系统,需要用到计算机技术、电子信息技术等
不同方面的知识,对维修技术人员自身的工作水平和能力
提出更高的要求
[1-4]
。本文主要简介GE Discovery XR656
DR系统硬件及全景拼接功能,并分享了一例全景拼接功能
故障的分析及排除过程,供同行参考。
系统”。现在普遍应用的DR主要是采用平板探测口对X
线产生的图像信号进行扫描和直接读出,成像原理是先将
X线信号转变为可见光通过利用硒层或碘化铯-光电2极
管组成的藻膜层进行聚集,由专门的读出电路直接读出直
接把X线光子转换成模拟电压供数字化后送计算机系统
进行处理,所有过程全部在平板探测器内完成。Discovery
XR656 DR系统由多个主要组件和子组件构成。如图1所
示系统主要组件有采集工作站、射线管高位托架(OTS)、
球管、系统柜(含高压发生器)、数字化探测器(两套GE
Flash Pad)、数字化扫描床和墙架。
1 Discovery XR656 DR系统硬件及全景拼接功能概述
1.1 Discovery XR656 DR系统硬件概述
DR是Digital radiography 的简称,即“数字放射成像
收稿日期:2018-12-17
作者邮箱:zxk1970@
图1 Discovery XR656 DR主要系统组件
采集工作站配备有专用计算机和图像数据库及放射控
制接口模块(Radiation Control Interface Module,RCIM)
中国医疗设备 2019年第34卷 10期
VOL.34 No.10
189
设备维修
DEVICE MAINTENANCE
和控制台手控开关,使用控制台手控开关可进行曝光,
RCIM主要用于控制系统的电源和复位功能。采集工作站
主要进行图像采集,图像显示和操作并通过DICOM标准
传输到其他工作站。射线管高位托架(OTS)是用来支撑
X射线管和OTS用户界面的定位设备。OTS组件可以在多
个方向上移动,以便在患者检查过程中正确进行装置定位。
系统柜中主要包含高压油箱、电源分配单元及各种通讯板
等。主要负责控制台、射线管高位托架OTS、扫描床、墙
架的电源分配及其相互间通讯。数字化探测器构造采用碳
纤维主要部件有面板和电子装置,负责原始数据的采集并
模数转换为数字图像数据,然后内部网络将此数据传输到
系统工作站,其他数字化墙架及扫描床用于完成各种放射
摄影程序的患者体位摆放与数字化探测器位置移动
[5-8]
。
1.2 全景拼接功能概述
全景拼接功能是一个高级应用程序,可将一个患者相
邻部位的两次甚至三次四次不同曝光区域的影像进行自动
拼接,从而突破了探测器尺寸的限制,可获得胸腹部、上
下肢等的整体影像。主要用于骨关节系统疾病等方面,能
可靠、准确地显示病变局部改变和全脊柱/全肢体的整体
受力状态,为临床诊断、术前方案的制定和术后疗效评估
等提供可靠依据
[9]
。自动图像粘贴可以在墙架和扫描床接
收器上使用。DR拼接图像的原始图像采集方式包括两种:
一种是球管与DR平板探测器做垂直上下的同步移动,另
一种是如图2a所示,X线球管组件相对静止在一个感兴趣
区中心位置,当DR平板探测器按箭头图示方向在做垂直
运动的同时,X线球管配合按箭头所示做上下转动角度。
系统先按顺序采集2~5次曝光采集到每个部位的影像,然
后再用数字方式将它们粘贴在一起,形成一张图像,采集
后的子图像会出现在机器浏览器屏幕上,而全景拼接的图
像会在大约10 s后出现
[10-11]
。
图2 墙架侧全景拼接采集示意图
注:a. 全景拼接硬件运动示意图;b. 每个部位影像;c. 全景
拼接图像。
2 Discovery XR656 DR设备全景拼接故障及其维修
方法
2.1 故障描述
设备开机后,设备信息提示栏偶发提示“system
control communication interrupted. Could be possible cable
disconnection. Call service if problem persists.”正常DR图
像功能正常,但频发全景拼接功能异常,提示“Exposure
中国医疗设备 2019年第34卷 10期
VOL.34 No.10
was prematurely ended. Image quality may be affected”等信
息,如图3所示,将图3a测试物品置于全景拼接患者固定
架上,反复功能测试,连续曝光频发中断(如图3b浏览器
图像缺失),全景拼接失败,无法正常连续曝光采集到如图
3c所示全景拼接成功获取图像,多次反复曝光捕捉故障点,
发现检查床及墙架侧进行全景扫描故障都会发生,并且与
球管及探测器的位置无关。
图3 全景拼接功能故障演示
注:a. 测试物品;b. 浏览器图像缺失;c. 全景拼接成功获取
图像。
2.2 故障分析
系统控制台信息提示栏偶发系统控制通讯中断提示,
但正常DR检查功能正常,说明正常DR曝光时,数据是
正常可以从平板探测器传至控制台,通讯正常,考虑到正
常DR曝光时间周期极短,说明系统内部通讯状态不稳定
而并未完全中断。而结合全景拼接功能故障时系统控制台
信息提示栏提示的曝光意外终止,综合分析全景拼接功能
故障可能的原因可以从以下几个方面去寻找:① 线路物理
连接问题导致的通讯故障,设备开机后,报通讯错误,最
大的可能是通讯连接线有松动,或者是插口处长时间使用
出现氧化,导致接触不良的虚接现象;② 软件问题,曝光
协议,应用程序故障,系统软件问题等;③ 相关硬件问题,
通信处理板、曝光控制盒等隐性硬件故障。
本着从低值到高值原则。首先检查负责通信的连接线
路是否正常及连接线接口是否因为机械运动等原因导致有
松动,或者是插口处长时间使用出现氧化,导致接触不良
的虚接等现象,其次是相关通讯板、电源板等模块的输入
输出特性,再次是排除曝光协议,应用程序故障,系统软
件等软件问题,最后排除通信处理板、曝光控制盒等隐性
硬件故障问题
[12-13]
。
2.3 故障处理
2.3.1 基本物理线路故障排查
关机断电状态下,进行基本物理线路故障排查,主要
发现以下异常点,第一处为如图4a所示,打开墙架前盖板
发现一通信线缆机械磨损严重,线标为“MS020021 FROM
WS A1J1 TO SKL(SYSTEMS CABINET)A25J14”,近墙架
端一处通信接口机械松动严重,使用扎带把线缆重新加固,
使其远离伸展臂运动轨道,并加固通信接口。第二处为如
图4b 所示系统柜内一通信接口松动,予以加固,关机重启,
予以全景拼接故障测试,故障依旧。
190
图4 物理线路异常点
注:a. 墙架物理线路异常点;b. 系统柜物理线路异常点。
2.3.2 软件故障排查
本着从简到繁原则,可先恢复曝光协议,若故障依旧
可重装系统及应用程序软件,若重装系统及应用程序软件
仍不能排除故障,排除系统及应用程序软件问题。当然在
软件故障排除的过程中需要注意的是在曝光协议及系统应
用程序进行还原前,应首先要确保所备份的最近日期的还
原协议及系统应用程序数据完好性,并且最好在还原前做
好备份,以保证万一不是软件问题所引起故障的情况时好
将数据及时还原回去,以便于下一步故障的分析判断。
首先是曝光协议故障排除,还原曝光协议,还原之
前确认最近日期的曝光协议管理备份)数据完好并做
好当前日期的曝光协议备份,Service模式下点击进入
System→Preferences→Protocols选项下有曝光协议的还原
与备份选项以供操作,使用最近日期的曝光协议备份数据
进行曝光协议还原,还原完成后,关机重启,予以全景拼
接故障测试,故障依旧,排除曝光协议导致的故障,使用
当前日期的曝光协议备份数据再次进行还原操作,以便机
器状态是故障排除前状态,供下一步的故障排除。
其次是系统及应用程序软件故障的排除,排除掉曝光
协议问题后,只能重装软件尝试。重装软件之前,确认最
近日期的系统备份数据完好并做好当前日期的系统备份,
若判断故障并非系统及应用程序软件故障引起,及时将设
备数据还原至最初故障排除前状态,供下一步的故障排除。
具体重装软件步骤如下。
(1)操作系统软件安装,将linux操作系统软件安装
光盘放入光驱,按住RESET 5秒钟,重启机器,在显示
器开始闪烁的时候,按住F9,之后会出现一个选择界面,
选择Optical Drive,点击回车,当出现Boot的时候,输出
GEHCXR(必须大写),回车,之后会自动完成安装,完成
安装后,取出光盘,系统会自动重启,进入linux系统。
(2)系统应用程序安装system applications software,
将最近日期的系统应用程序软件安装光盘放入光驱,点击
CDROM图标,并且点击并且运行autorun,机器会自动重
启,当机器重启后,再次打开CDROM图标,并再次运行
autorun,在出现的对话框中选择相对应的机型656,系统
会自动完成安装,当完成安装后,机器会重启,重启后取
出光盘。
DEVICE MAINTENANCE
设备维修
(3)重启完成后,输入账户名:insite密码:2getin进
入应用操作平台,点击右上角的工具图标,选择维修图标,
在维修界面选择Utillties并找到System,找到Restore,将
最近日期的系统备份放入光驱,选择从光驱还原,点击开始,
此处要注意,在如图5所示,出现的对话框中,一定要一
个一个的完成还原,此处应注意,在还原下一选项前一定
要确保上一选项的还原进程有“success”提示出现,在完
成所有选项的还原后,点击退出,将光盘取出。
图5 系统还原弹窗
(4)将曝光协议光盘放入光驱,进入参数设置界面,
选择参数-协议-还原,完成协议的还原,还原完成后,
重启设备,测试PACS和WORKLIST的功能正常,全景拼
接功能测试,故障依旧,且通信报错仍存在,此时应将故
障前所做系统备份数据及曝光协议管理备份数据进行还原,
以便机器状态是故障排除前状态,供下一步的故障排除。
2.3.3 硬件故障排查
鉴于“多次反复曝光捕捉故障点,发现检查床及墙架
时进行全景扫描故障都会发生,且尚未发现故障与球管及
探测器的某一具体位置点关联”的故障现象,可重点考虑
关注检查床及墙架共用部分设备部件如高压柜与控制台部
分,结合控制台一直存在通信故障报错,最后综合分析故
障主要涉及硬件部分优先从通信部分排查,而涉及通信的
物理硬件主要有图6所示系统主通信板和图7所示RCIM。
本着由简入繁原则,先从主通信板故障排查,关机状态下,
打开机箱,负责通信的板子在供电柜门上,将通信板外部
的保护外壳拆下, 各连接线予以重新拔插,无松动,开机
观察通信板上常规的指示灯,指示灯的工作状态都正常。
此时在开机上电状态下用数字万用表测量通信板开关电源
供电电压,电压值为DC23 V,调节如图所示开关电源点位
器,将输出调至24 V,关机重启,功能测试,设备故障依
旧,基本排除通讯板故障,最后只能怀疑图7所示RCIM,
检查其工作状态指示灯及供电电源未见异常,考虑到其板
级集成度极高,只能考虑将其更换,更换RCIM后,系统
通信报错提示消失,全景拼接功能多次测试正常,曝光连
续不中断,能够获取采集到图3c图像所示的图像。
中国医疗设备 2019年第34卷 10期
VOL.34 No.10
191
设备维修
DEVICE MAINTENANCE
[参考文献]
[1] 杨小娟,吴国荣.医疗设备维修管理现状及对策探讨[J].中国
医疗器械信息,2017,74(18):152-153.
[2] 范境文.医院医疗设备维修精细化管理的必要性和措施[J].医
疗装备,2017,30(18):73-74.
[3] 辛胜昌.浅论本院医疗设备维修管理中存在的问题及对策[J].
中国医疗器械信息,2017,23(16):149-151.
[4] 陈晓飞,周晟,汪新柱,等.数字X线摄影焦点至探测器
图6 系统主通信板
距离与体模表面入射剂量的实验研究[J].中国临床研
究,2016,29(3):384-385.
[5] 韩丰谈.医学影像设备学[M].北京:人民卫生出版社,2004:458-
图7 放射控制接口模块(RCIM)
464.
[6] 徐江波.GE Definium6000飞天 DR的使用管理与常见故障维
修[J].医疗装备,2018,31(3):151.
[7] 谈伟,张德洲,黄林,等.数字X线摄影的全景拼接技术在骨关
节系统中的应用研究[J].华西医学,2013,(9):1398-1401.
[8] 熊刚,朱俊义,魏瑜栋.新形势下我院医疗设备维修管理模式
改进方法探讨[J].中国医疗设备,2018,33(9):175-177.
[9] 徐跃,梁碧玲.医学影像设备学[M].3版.北京:人民卫生出版
社,2010:257.
[10] 向思伟,郑琨.GE Defnium6000 DR 故障维修两例[J].中国医
疗设备,2018,33(2):100-101.
[11] 刘伟宾,张燕,吴新淮,等.GE Discovery XR650 X线系统
在儿童双下肢不等长中的临床应用价值[J].医疗卫生装
备,2013,34(8):69-70.
[12] 盛况,孔德友,范磊.GE Defnium 6000数字X线机故障维修分
析3例[J].中国医疗设备,2018,33(1):102-103.
设备,2016,31(10):172-173.
械信息,2017,23(19):130-131.
[13] 于金科,祝健.Defnium 6000 DR 故障维修3例[J].中国医疗
[14] 苏健.加强医疗设备维修服务管理的新策略[J].中国医疗器
[15] 王可寿.医院医疗设备维修的精细化管理[J].医疗装备,
2018,31(9):93-94.
C
3 总结
本文简介了GE Discovery XR656 DR系统硬件及全景
成像功能,并分享了一例全景拼接功能故障的故障现象,
并给出了故障分析及排除过程。随着微电子等各种高新技
术在医疗设备中广泛运用与整合,加之医疗设备厂方的技
术垄断与保护,使得大型医疗设备的维修和维护面临严峻
考验,医院工程师传统的维修模式面临巨大挑战。作为院
方工程师,应从设备装机安装调试开始就应积极加强自身
对机器的了解学习,与此同时,当初期设备在保修期内时,
应加强与厂家工程师的沟通与学习,从机器的基本操作使
用到基本结构原理,从巡检保养各项明细到日常常见故障
的分析的思路和排除方法,为以后独立维修打下良好基础。
而在设备的日常管理当中,应加强设备的巡检保养,制定
严格的巡检保养明细和周期,按期执行并落实到每条明细,
将传统的设备故障后维修的维修模式升级为预防性维修,
及时将故障隐患排除,并且在每次故障后及时详细的记录
故障现象、分析解决的全过程,做好每台设备的动态管理
档案,给设备的精细化、信息化管理提供数据支持。各级
医院应巧妙地运用精细化管理模式,高度重视维修人员专
业技能的培养,提高维修设备的信息化,以此来保障医疗
设备能够高效、正常的投入临床运用中