最新消息: USBMI致力于为网友们分享Windows、安卓、IOS等主流手机系统相关的资讯以及评测、同时提供相关教程、应用、软件下载等服务。

追踪者3200s和4.3软件入门指南说明书

IT圈 admin 53浏览 0评论

2024年5月11日发(作者:瞿巧兰)

Getting Started instructions for a Tracker 3200S using

the Huntron Workstation 4.3 software

1-1 Recommended Controller (PC)

The following are provided to assist you in purchasing the correct computer system to use with the

Huntron Workstation software.

• Multi-core processor; Intel i7 or similar recommended

• Microsoft Windows 10

• 8GB+ RAM

• Solid state hard drive

• 1920x1080 video resolution

• DVD drive

• Keyboard and Mouse

• Internet Connection

• One available USB port

1-2 Installing the Huntron Workstation Software

NOTE: Install the software before connecting the Tracker 3200S to the PC.

1.

Power up your computer. Insert the software DVD into the optical drive.

2.

If your drive was configured for Auto-run, a menu will be displayed. Click the

WorkstationSetup item. If no menu is displayed, run the

program located on the DVD.

3.

A Huntron License Agreement form is displayed. If you agree, select the “I accept

the license agreement” option. Click Next. The file on the DVD will

be displayed. Click Next

4.

The default destination folder is C:Program FilesHuntron Workstation. Click

Next if you do not want to change the default folder.

5.

Select the options for desktop and taskbar icons.

6.

Click Next to begin installation.

7.

If you will not be connecting your Tracker 3200S to an Access Prober, answer NO

to the questions regarding Access Prober related Camera and NI Vision

installations.

8.

Click Finish when the “Huntron Workstation has been successfully installed”

window is displayed.

Getting started instructions for Tracker 3200S and 4.3 Software

21-2615 Version 1.2

1

Do not run the software at this point. The hardware must now be installed to

complete the installation.

1-3 CONNECTING THE TRACKER 3200S

Note for Windows 64bit users: The device driver may attempt to auto install and fail.

After following the steps below (they may not all occur), open Device Manager and look

under Huntron devices for the Tracker 3200S with a yellow exclamation point. Update this

driver by right clicking on them. Select the install DVD for the source of the driver if

needed.

1. Connect the Tracker 3200S power cord to a power outlet.

2. Connect the USB cable to the Tracker 3200S and the PC.

3. Turn on the Tracker 3200S power

4. An Add New Hardware wizard prompts you to install a “Huntron Tracker Model

3200S”. If prompted to find driver on the Internet select No and Click Next.

5. If the wizard prompts you to automatically find drivers, click Next.

6. When a message appears, indicating that the driver has not passed Windows Logo

testing, click “Continue anyway”. Then click Finish.

7. Install any other equipment to be used with the Huntron Workstation software.

8. Remove your Huntron Workstation DVD. You are now ready to use the software.

9. A Huntron Workstation icon was created on your windows desktop. Double click

this icon to start the software.

10. The Huntron Workstation Activation dialog is displayed. This software is only

licensed to be used on one Tracker 3200S. The activation codes are generated by

Huntron and are specific to the Tracker 3200S serial number. Activation codes are

printed on the software DVD. If you do not know the activation codes, call Huntron

technical support (425-743-3171) and provide the serial number for the Tracker

3200S. Tracker 3200S serial numbers start with a 32S.

1-4

1.

CONFIGURING THE TRACKER 3200S

When the Workstation main screen is displayed, on the main menu, select Tools-

>Options->Hardware. Set the Tracker to Tracker 3200S. Set the Tracker Port to

USB.

Set the Prober settings as needed (refer to Prober Getting Started document).

Click the connect buttons for the Tracker and Prober if used.

Click the Save button.

NOTE: The Tracker 3200S is already calibrated at the factory. However, if you are

connecting a Tracker 3200S to a Prober, you will need to calibrate it as a complete

system.

Connect the Tracker 3200S to your Prober, follow the steps below to calibrate the

system.

Tracker 3200S Calibration for Prober

2.

3.

4.

5.

6.

7.

To access the calibration window, click Tools->Maintenance->Tracker3200S

Diagnostics.

Getting started instructions for Tracker 3200S and 4.3 Software

21-2615 Version 1.2

2

Do not click the Verify Hardware button. This routine requires special hardware and

test procedures.

Click the Alignment button. Ensure that all cables are connected. The calibration routine

will run for a few minutes. Do not power off any of the instruments during this process.

After a successful calibration, click the Verify Alignment button. If you get a “successful

verification”, click Close to exit the Diagnostic window. If you get a “Verification Failed”

message, Re-run the Alignment routine and run the Verify Alignment again. If it still

fails, call Huntron Technical support.

Return to Tools->Options->Hardware and click the Connect button(s) for the appropriate

hardware.

1-5 Follow the software tutorial.

A software tutorial is included in the software DVD Documents folder and in the default

installed folder (C:Program FilesHuntron Workstation 4.3Documents). The tutorial

can also be accessed via the Help menu in the software. If you purchased an Access

Prober follow the Workstation , otherwise follow the Workstation Tutorial

Tracker . The tutorial will enable a new user to get acquainted with most of the

software features. The help menu in the software will also assist you.

Whenever you need detailed information, click Help in the appropriate screens.

All user’s manuals are located in the Manual CD (06-5217) included in your package.

Use the Tracker 3200S file to view or print the manual.

If you need further assistance, call the Huntron Technical support at

1-(800)-426-9265 or (425)-743-3171

email:****************

--------end of document--------

Getting started instructions for Tracker 3200S and 4.3 Software

21-2615 Version 1.2

3

2024年5月11日发(作者:瞿巧兰)

Getting Started instructions for a Tracker 3200S using

the Huntron Workstation 4.3 software

1-1 Recommended Controller (PC)

The following are provided to assist you in purchasing the correct computer system to use with the

Huntron Workstation software.

• Multi-core processor; Intel i7 or similar recommended

• Microsoft Windows 10

• 8GB+ RAM

• Solid state hard drive

• 1920x1080 video resolution

• DVD drive

• Keyboard and Mouse

• Internet Connection

• One available USB port

1-2 Installing the Huntron Workstation Software

NOTE: Install the software before connecting the Tracker 3200S to the PC.

1.

Power up your computer. Insert the software DVD into the optical drive.

2.

If your drive was configured for Auto-run, a menu will be displayed. Click the

WorkstationSetup item. If no menu is displayed, run the

program located on the DVD.

3.

A Huntron License Agreement form is displayed. If you agree, select the “I accept

the license agreement” option. Click Next. The file on the DVD will

be displayed. Click Next

4.

The default destination folder is C:Program FilesHuntron Workstation. Click

Next if you do not want to change the default folder.

5.

Select the options for desktop and taskbar icons.

6.

Click Next to begin installation.

7.

If you will not be connecting your Tracker 3200S to an Access Prober, answer NO

to the questions regarding Access Prober related Camera and NI Vision

installations.

8.

Click Finish when the “Huntron Workstation has been successfully installed”

window is displayed.

Getting started instructions for Tracker 3200S and 4.3 Software

21-2615 Version 1.2

1

Do not run the software at this point. The hardware must now be installed to

complete the installation.

1-3 CONNECTING THE TRACKER 3200S

Note for Windows 64bit users: The device driver may attempt to auto install and fail.

After following the steps below (they may not all occur), open Device Manager and look

under Huntron devices for the Tracker 3200S with a yellow exclamation point. Update this

driver by right clicking on them. Select the install DVD for the source of the driver if

needed.

1. Connect the Tracker 3200S power cord to a power outlet.

2. Connect the USB cable to the Tracker 3200S and the PC.

3. Turn on the Tracker 3200S power

4. An Add New Hardware wizard prompts you to install a “Huntron Tracker Model

3200S”. If prompted to find driver on the Internet select No and Click Next.

5. If the wizard prompts you to automatically find drivers, click Next.

6. When a message appears, indicating that the driver has not passed Windows Logo

testing, click “Continue anyway”. Then click Finish.

7. Install any other equipment to be used with the Huntron Workstation software.

8. Remove your Huntron Workstation DVD. You are now ready to use the software.

9. A Huntron Workstation icon was created on your windows desktop. Double click

this icon to start the software.

10. The Huntron Workstation Activation dialog is displayed. This software is only

licensed to be used on one Tracker 3200S. The activation codes are generated by

Huntron and are specific to the Tracker 3200S serial number. Activation codes are

printed on the software DVD. If you do not know the activation codes, call Huntron

technical support (425-743-3171) and provide the serial number for the Tracker

3200S. Tracker 3200S serial numbers start with a 32S.

1-4

1.

CONFIGURING THE TRACKER 3200S

When the Workstation main screen is displayed, on the main menu, select Tools-

>Options->Hardware. Set the Tracker to Tracker 3200S. Set the Tracker Port to

USB.

Set the Prober settings as needed (refer to Prober Getting Started document).

Click the connect buttons for the Tracker and Prober if used.

Click the Save button.

NOTE: The Tracker 3200S is already calibrated at the factory. However, if you are

connecting a Tracker 3200S to a Prober, you will need to calibrate it as a complete

system.

Connect the Tracker 3200S to your Prober, follow the steps below to calibrate the

system.

Tracker 3200S Calibration for Prober

2.

3.

4.

5.

6.

7.

To access the calibration window, click Tools->Maintenance->Tracker3200S

Diagnostics.

Getting started instructions for Tracker 3200S and 4.3 Software

21-2615 Version 1.2

2

Do not click the Verify Hardware button. This routine requires special hardware and

test procedures.

Click the Alignment button. Ensure that all cables are connected. The calibration routine

will run for a few minutes. Do not power off any of the instruments during this process.

After a successful calibration, click the Verify Alignment button. If you get a “successful

verification”, click Close to exit the Diagnostic window. If you get a “Verification Failed”

message, Re-run the Alignment routine and run the Verify Alignment again. If it still

fails, call Huntron Technical support.

Return to Tools->Options->Hardware and click the Connect button(s) for the appropriate

hardware.

1-5 Follow the software tutorial.

A software tutorial is included in the software DVD Documents folder and in the default

installed folder (C:Program FilesHuntron Workstation 4.3Documents). The tutorial

can also be accessed via the Help menu in the software. If you purchased an Access

Prober follow the Workstation , otherwise follow the Workstation Tutorial

Tracker . The tutorial will enable a new user to get acquainted with most of the

software features. The help menu in the software will also assist you.

Whenever you need detailed information, click Help in the appropriate screens.

All user’s manuals are located in the Manual CD (06-5217) included in your package.

Use the Tracker 3200S file to view or print the manual.

If you need further assistance, call the Huntron Technical support at

1-(800)-426-9265 or (425)-743-3171

email:****************

--------end of document--------

Getting started instructions for Tracker 3200S and 4.3 Software

21-2615 Version 1.2

3

与本文相关的文章

发布评论

评论列表 (0)

  1. 暂无评论