Manual Supplement
Manual Title: 5320A Users
Print Date: September 2006
Revision/Date:
Supplement Issue: 2
Issue Date: 11/06
3
Page Count:
This supplement contains information necessary to ensure the accuracy of the
above manual. This manual is distributed as an electronic manual on the
following CD-ROM:
CD Title: 5320A
CD Rev. & Date: 9/2006
PCD N: 2634346
© 2006 Fluke Corporation. All rights reserved. Printed in U.S.A.
x
5320A Users Manual Supplement
Change #1
On page 7-26, replace Figure 7-23 with the following:
Quadtech Sentry
Fluke 5320A
FrontBack
Fluke 5320A kV Divider/R Multiplier
Figure 7-23. 10 kV High Voltage Adapter Application
Change #2
On page 5-15, change the following:
From:
OUTPute[:STATus](?)
OUTPut[:STATe](?)
To:
10/06 1
Manual Supplement 5320A Users
Change #3
On page 5-25, change the following:
From: [SOUR]:SAF:IDP[:CURR]:RES?
Description: This command returns the measured current flowing through the UUT. If not
already selected, it also switches the Calibrator to the Passive Leakage Current
function.
Query:
To:
SAF:IDP:RES? Returns the value of the instantaneous current flowing
through the Calibrator.
[SOUR]:SAF:IDP[:CURR]:RES?
Description: This command returns the measured current flowing through the UUT. If not
already selected, it also switches the Calibrator to the Passive Leakage Current
function.
Query: SAF:IDP:RES? Returns the value of the resulting measured UUT
current.
On page 5-26, change the following:
From: [SOUR]:SAF:IDD[:CURR]: RES?
Description: This command returns the measured current flowing through the UUT. If not
already selected, it also switches the Calibrator to the Differential Leakage
Current function.
Query:
To:
SAF:IDD:RES? Returns the value of the instantaneous current flowing
through the Calibrator.
[SOUR]:SAF:IDD[:CURR]: RES?
Description: This command returns the measured current flowing through the UUT. If not
already selected, it also switches the Calibrator to the Differential Leakage
Current function.
Query: SAF:IDD:RES? Returns the value of the resulting measured UUT
current.
Change #4
On page 6-6, after step 2 of High Resistance Source Verification, add the following note:
Note
For some megohmmeters, when using the Calibrator’s 100 G
Ω
value or when using the
resistance multiplier adapter, the leads must be swapped between the Calibrator’s HI
and LO ohms resistance output. The ground must be turned on (GDN ON) when
swapping HI and LO lead positions in the high ohms resistance function. For example to
make a proper measurement with the Quadtech 1865 megohmmeter, connect the HI
terminal on the megohmmeter to the LO terminal on the Calibrator and connect the LO
terminal on the megohmmeter to the HI terminal on the Calibrator. Turn the ground on
and proceed to make the measurement.
2 10/06
5320A Users Manual Supplement
Change #5
On page 1-8, delete the following from the Table:
Plug adapter (x2) 2743474
11/06 3
2024年5月21日发(作者:旁念)
Manual Supplement
Manual Title: 5320A Users
Print Date: September 2006
Revision/Date:
Supplement Issue: 2
Issue Date: 11/06
3
Page Count:
This supplement contains information necessary to ensure the accuracy of the
above manual. This manual is distributed as an electronic manual on the
following CD-ROM:
CD Title: 5320A
CD Rev. & Date: 9/2006
PCD N: 2634346
© 2006 Fluke Corporation. All rights reserved. Printed in U.S.A.
x
5320A Users Manual Supplement
Change #1
On page 7-26, replace Figure 7-23 with the following:
Quadtech Sentry
Fluke 5320A
FrontBack
Fluke 5320A kV Divider/R Multiplier
Figure 7-23. 10 kV High Voltage Adapter Application
Change #2
On page 5-15, change the following:
From:
OUTPute[:STATus](?)
OUTPut[:STATe](?)
To:
10/06 1
Manual Supplement 5320A Users
Change #3
On page 5-25, change the following:
From: [SOUR]:SAF:IDP[:CURR]:RES?
Description: This command returns the measured current flowing through the UUT. If not
already selected, it also switches the Calibrator to the Passive Leakage Current
function.
Query:
To:
SAF:IDP:RES? Returns the value of the instantaneous current flowing
through the Calibrator.
[SOUR]:SAF:IDP[:CURR]:RES?
Description: This command returns the measured current flowing through the UUT. If not
already selected, it also switches the Calibrator to the Passive Leakage Current
function.
Query: SAF:IDP:RES? Returns the value of the resulting measured UUT
current.
On page 5-26, change the following:
From: [SOUR]:SAF:IDD[:CURR]: RES?
Description: This command returns the measured current flowing through the UUT. If not
already selected, it also switches the Calibrator to the Differential Leakage
Current function.
Query:
To:
SAF:IDD:RES? Returns the value of the instantaneous current flowing
through the Calibrator.
[SOUR]:SAF:IDD[:CURR]: RES?
Description: This command returns the measured current flowing through the UUT. If not
already selected, it also switches the Calibrator to the Differential Leakage
Current function.
Query: SAF:IDD:RES? Returns the value of the resulting measured UUT
current.
Change #4
On page 6-6, after step 2 of High Resistance Source Verification, add the following note:
Note
For some megohmmeters, when using the Calibrator’s 100 G
Ω
value or when using the
resistance multiplier adapter, the leads must be swapped between the Calibrator’s HI
and LO ohms resistance output. The ground must be turned on (GDN ON) when
swapping HI and LO lead positions in the high ohms resistance function. For example to
make a proper measurement with the Quadtech 1865 megohmmeter, connect the HI
terminal on the megohmmeter to the LO terminal on the Calibrator and connect the LO
terminal on the megohmmeter to the HI terminal on the Calibrator. Turn the ground on
and proceed to make the measurement.
2 10/06
5320A Users Manual Supplement
Change #5
On page 1-8, delete the following from the Table:
Plug adapter (x2) 2743474
11/06 3